一、電導率儀的電極常數
電(dian)(dian)導率儀的電(dian)(dian)極(ji)(ji)常數(shu)是指電(dian)(dian)導電(dian)(dian)極(ji)(ji)的長度(du)與面(mian)積之間(jian)的比(bi)值,這個比(bi)值影響電(dian)(dian)極(ji)(ji)的靈(ling)敏度(du)。電(dian)(dian)極(ji)(ji)常數(shu)越小(xiao),意味著電(dian)(dian)極(ji)(ji)面(mian)積越大,從(cong)而使得電(dian)(dian)極(ji)(ji)的靈(ling)敏度(du)更高。
電(dian)極常(chang)(chang)數(shu)可以通過(guo)公式K=1/(σ*A/D)計算得(de)出,其中K是電(dian)極常(chang)(chang)數(shu)(單(dan)位(wei):S/cm),σ是介質的電(dian)導(dao)(dao)率(lv)(單(dan)位(wei):S/cm),A是電(dian)極面積(單(dan)位(wei):cm2),D是電(dian)極間(jian)距(單(dan)位(wei):cm)。此外,電(dian)極常(chang)(chang)數(shu)也可以通過(guo)在特定(ding)濃(nong)度(du)的KCL溶(rong)液中測量電(dian)導(dao)(dao)電(dian)極的電(dian)導(dao)(dao)來測定(ding),此時KCL溶(rong)液的電(dian)導(dao)(dao)率(lv)是已知(zhi)的。
常(chang)(chang)見的電(dian)導(dao)電(dian)極常(chang)(chang)數有0.01、0.1、1.0、10四種類(lei)型,具(ju)體值(zhi)通常(chang)(chang)標(biao)記在(zai)電(dian)導(dao)電(dian)極上(shang),用戶可以根據電(dian)極上(shang)標(biao)注的常(chang)(chang)數來調節儀器面板上(shang)的“常(chang)(chang)數”補償調節旋鈕,以確(que)保測(ce)量結果的準確(que)性(xing)。
二、電導率儀的單位介紹
電(dian)導率儀(yi)的單位(wei)通(tong)常是西門子/米(S/m)。
西(xi)門(men)子/米是(shi)電(dian)阻(zu)率(ρ)的(de)倒數(shu),其中σ=1/ρ。換句話(hua)說,當(dang)1安培(1A)的(de)電(dian)流通過物體的(de)橫截面,并在存(cun)在1伏(fu)特(te)(1V)的(de)電(dian)壓時(shi),該物體的(de)電(dian)導就是(shi)1S。西(xi)門(men)子實際上(shang)等效于(yu)1安培/伏(fu)特(te)。
在實際應用中,也(ye)常用毫西門子(zi)(mS/m)和微西門子(zi)(μS/cm)作為單(dan)位。例(li)如,1μS/cm=0.001mS/m;1000μS/cm=1mS/m。這些單(dan)位的使(shi)用取決于具體的測量需求和被測物質的類型。